We evaluate the performance of a rapidly responsive adaptive algorithm (VDL1.1) for automated oxygen control in preterm infants with respiratory insufficiency.
Design Interventional cross-over study of a 24-hour period of automated oxygen control compared with aggregated data from two flanking periods of manual control (12 hours each).
Participants Preterm infants receiving non-invasive respiratory support and supplemental oxygen; median birth gestation 27 weeks (IQR 26–28) and postnatal age 17 (12–23) days.
Intervention Automated oxygen titration with the VDL1.1 algorithm, with the incoming SpO2 signal derived from a standard oximetry probe, and the computed inspired oxygen concentration (FiO2) adjustments actuated by a motorised blender.
The desired SpO2 range was 90%–94%, with bedside clinicians able to make corrective manual FiO2 adjustments at all times.